We import your parts

ClimMate Set aw

ClimMate Set aw

With the large sample chamber, larger sample quantities or pieces can also be measured. No temperature control and no humidity balance monitoring. This instrument is suitable for the first AW measured value and can be used as a portable version.

Measurement Specifications

 

Parameter

Water Activity

Temperature

Measurement Principle

Resistive Electrolytic

Surface-IR and NTC

Measurement Range

0.03……1.00aw

0…50ºC (32…122ºF)

Calibration Range

0.11……0.90aw

N/A

Resolution

0.01aw

0.1ºC

Accuracy

±0.02aw within cal. range

±0.15°C

Precision

±0.01aw within cal. range*

±0.1°C

Repeatability

±0.01aw within cal. range *

N/A

Temperature Control

No temperature control and humidity equilibrium detection available

Calibration Points (%rH)

Only with nSoft-CAL:

Points: 11%, 33%, 53%, 75%, 90%

Instrument-Specifications

 

ClimMate Set aw

Versions

N/A

Current Supply

4x1,5 VDC Alkaline batteries or adapter 5 VDC

Power Supply/Mains

90…264 VAC 50/60 Hz

Display

Capacitive TFT display

Communication

USB to PC

Housing

- Plastic Housing, Weight: 0.2kg

- Dimensions (L x W x H) = 62 x 170 x 34mm

Volume measurement chamber

480 mL

Dimensions Sample Cup

Diameter 80mm x Height 95mm

Protection Class

IP54

21CFR11

Not Compliant