GMP5F RF
-80ºC to + 140ºC
Applications: Autoclaves, incubators, ovens, on-site steam, stability chambers, warehouses, lyophilisates, freezers, ultra-freezers, refrigerators, cold rooms, cold chain shipments, logistics.
Specs:
· Collects and transmits temperature in real time to XpertLog software
· ITS-90 calibration coefficients are stored in internal memory
· Complies with FDA 21 CFR Part 11
· Network ready: multiple users, centralized, server database, active directory authentication
· Can be activated remotely
Data sheet
Characteristics |
GMP5F-RF |
|
Sensor type |
PT-100 |
|
- flexible |
480 mm (19”) |
|
- customized |
up to 5000 mm (195”) |
|
- sensor size |
1.5 x 8 mm (0.06 x 0.315”) |
|
- cross section |
1.77 mm² (0.0027-inch²) |
|
- body |
-80°C to +140°C |
|
- sensor |
-200°C to +200°C |
|
Accuracy |
±0.15°C from -80 to +140°C |
|
Resolution |
0.01°C |
|
Memory capacity |
32,000 data points / 6400 data points per channel |
|
Memory type |
nonvolatile |
|
Sampling rate |
1 sec to 24 hours |
|
Transmission |
simultaneous:
maximum 20 loggers |
|
Internal clock drift |
4 sec / 24 h @ 23°C |
|
Mechanical |
||
- material |
Peek 1000 - FDA compliant |
|
- enclosure |
IP-68 |
|
- weight |
170 g (6.00 oz) |
|
- diameter |
65 mm (2.56”) |
|
- height |
55 mm (2.17") |
|
Power |
2 x Lithium batteries,
0.35 g/each Lithium content |
|
Battery lifetime |
Up to 6 months, based on usage |
|
Antenna |
Internal |
|
Wireless range |
121 m (400 feet) |
|
Radio frequency |
2.4 Ghz ISM band |
|
Access point |
USB to computer |
|
Calibrations |
Factory
Calibrations: User Calibrations: |
traceable
NIST/COFRAC - ITS-90 coefficients stored in the internal memory |
Standard Calibration Points |
• -80°C |
|
Certifications |
EMC: |
EN 60601-1-2 / 2007 |