We import your parts

GMPF 10R SS

GMPF 10R SS

Small size real-time temperature data logger with flexible temperature sensor, to be placed between vials. It adapts to the height and diameter of 10R vials (-60ºC to + 85ºC).

-60ºC to + 85ºC

Applications: Autoclaves, incubators, ovens, on-site steam, stability chambers, warehouses, lyophilisates, freezers, ultra-freezers, refrigerators, cold rooms, cold chain shipments, logistics.

Specs:

· Collects and transmits temperature in real time to XpertLog software.

· ITS-90 calibration coefficients are stored in internal memory.

· Complies with FDA 21 CFR Part 11

· Network ready: multiple users, centralized, server database, active directory authentication

· Calibration can be done on site.

· Can be activated remotely

Data sheet

Characteristics

GMPF_RF_10R_SS_THIN

Sensor type

4 wires thin

- flexible

150 mm (6”) or custom length

- sensor size

1.2 x 2.8 mm (0.05 x 0.11”)

- cross section

1.13 mm² (0.00175-inch²)

- body

-60°C to +85°C

- sensor

-80°C to +140°C

Accuracy

±0.1°C from -60 to +85°C

Resolution

0.01°C

Memory capacity

32,000 data points

Memory type

nonvolatile

Sampling rate

1 sec to 24 hours

Transmission

simultaneous: maximum 20 loggers
sequential: unlimited

Internal clock drift

4 sec / 24 h @ 23°C

Mechanical

- material

Peek 1000 - FDA compliant

- enclosure

IP-68

- weight

24 g (0.85 oz)

- diameter

25 mm (0.98”)

- height

46 mm (1.81”)

Power

1 x Lithium batteries, 0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970,
user replaceable

Battery lifetime

Up to 2 months, based on usage

Antenna

Internal

Wireless range

121 m (400 feet)

Radio frequency

2.4 Ghz ISM band

Access point

USB to computer
Programs and reads unlimited number of loggers

Calibrations

Factory Calibrations:

User Calibrations:

traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
closed loop calibration using XpertLog® software

Standard Calibration Points

• -60°C
• 0°C
• 85°C

Certifications

EMC:
Health Safety:
RF:

EN 60601-1-2 / 2007
EN 60950-1 / 2006
EN 300440-2 v1.3.1 (2009-03)