We import your parts

GMPF CO2 RF

GMPF CO2 RF

Real-time CO2, temperature and pressure data logger and sends data to XpertLog® software (0 to 20% CO2).

0 to 20% CO2

With an accuracy of ± 0.5% CO2 and ± 0.1 ° C, GMPF CO2 RF allows you to measure CO2 up to 20% and temperature from -200 to + 200 ° C.

Applications: Incubators, refrigerators, chain shipments of cold.

Specs:

· Collects and transmits CO2 and temperature in real time to XpertLog software

· ITS-90 calibration coefficients are stored in internal memory

· Complies with FDA 21 CFR Part 11

· Network ready: multiple users, centralized, server database, active directory authentication

· Can be activated remotely

Data sheet

Characteristics

GMPF-CO2_RF

Sensor type

CO2 - IR / Temp: 4 wires PT-100

- flexible

480 mm (19”)

- customized

up to 5000 mm (195”)

- sensor size

1.5 x 8 mm (0.06 x 0.315”)

- cross section

1.77 mm² (0.0027-inch²)

- body

0 to 20% CO2
0°C to 50°C
300 to 1100 mBar (built-in pressure compensation)

- sensor

-200°C to +200°C

Accuracy

±0.1°C from -90°C to +140°C
±0.5% CO2 from 0 to 10%CO2

Resolution

0.01% CO2 / 0.01°C

Memory capacity

16,000 data points CO2 / 16,000 data points temperature

Memory type

nonvolatile

Sampling rate

2 min to 24 hours

Transmission

simultaneous: maximum 20 loggers
sequential: unlimited

Internal clock drift

4 sec / 24 h @ 23°C

Mechanical

- material

Peek 1000 - FDA compliant

- enclosure

IP-30

- weight

153 g (5.39 oz)

- diameter

64.5 mm (2.53”)

- height

52 mm (2.04”)

Power

2 x Lithium batteries, 0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970,
user replaceable

Battery lifetime

Up to 6 months, based on usage

Antenna

Internal

Wireless range

121 m (400 feet)

Radio frequency

2.4 Ghz ISM band

Access point

USB to computer
Programs and reads unlimited number of loggers

Calibrations

Factory Calibrations:

User Calibrations:

traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
closed loop calibration using XpertLog® software

Standard Calibration Points

0°C
23°C
50°C

Certifications

EMC:
Health Safety:
RF:

EN 60601-1-2 / 2007
EN 60950-1 / 2006
EN 300440-2 v1.3.1 (2009-03)