-60ºC to + 85ºC
Applications: freeze dryers, freezers, refrigerators, cold rooms.
Specs:
· Collects and transmits temperature in real time to XpertLog software.
· ITS-90 calibration coefficients are stored in internal memory.
· Complies with FDA 21 CFR Part 11
· Network ready: multiple users, centralized, server database, active directory authentication
· Calibration can be done on site.
· Can be activated remotely
Data sheet
Characteristics |
GMPF_RF_Mini_THIN/GMPF_RF_Mini_THIN_1m |
|
Sensor type |
4 wires thin |
|
- flexible |
150 mm (6”) or custom length |
|
- sensor size |
1.2 x 2.8 mm (0.05 x 0.11”) |
|
- cross section |
1.13 mm² (0.00175-inch²) |
|
- body |
-60°C to +85°C |
|
- sensor |
-80°C to +140°C |
|
Accuracy |
±0.1°C from -60 to +85°C |
|
Resolution |
0.01°C |
|
Memory capacity |
32,000 data points |
|
Memory type |
nonvolatile |
|
Sampling rate |
1 sec to 24 hours |
|
Transmission |
simultaneous: maximum 20 loggers |
|
Internal clock drift |
4 sec / 24 h @ 23°C |
|
Mechanical |
||
- material |
Peek 1000 - FDA compliant |
|
- enclosure |
IP-68 |
|
- weight |
19 g (0.67 oz) |
|
- diameter |
22 mm (0.86”) |
|
- height |
56 mm (2.20”) |
|
Power |
1 x Lithium batteries, 0.35 g/each Lithium content |
|
Battery lifetime |
Up to 6 months, based on usage |
|
Antenna |
Internal |
|
Wireless range |
121 m (400 feet) |
|
Radio frequency |
2.4 Ghz ISM band |
|
Access point |
USB to computer |
|
Calibrations |
Factory Calibrations: User Calibrations: |
traceable NIST/COFRAC - ITS-90 coefficients stored
in the internal memory |
Standard Calibration Points |
• -60°C |
|
Certifications |
EMC: |
EN 60601-1-2 / 2007 |