We import your parts

GMPF P RF

GMPF P RF

Real-time pressure data logger to monitor pressure, with flexible temperature sensor (0.01 to 5 BarA).

0.01 to 5 BarA

Highly efficient (fast validation time), impressive precision (± 0.1 ° C).

Applications: autoclaves, incubators, ovens, on-site steam, stability chambers, warehouses, freezers, refrigerators, cold rooms, cold chain shipments, logistics.

Specs:

· Collects and transmits pressure and temperature in real time to XpertLog software

· ITS-90 calibration coefficients are stored in internal memory

· Complies with FDA 21 CFR Part 11

· Network ready: multiple users, centralized, server database, active directory authentication

· Can be activated remotely

Data sheet

Characteristics

GMPF-P_RF

Sensor type

Absolute pressure / Temp: 4 wires PT-100

- flexible

480 mm (19”)

- customized

up to 5000 mm (195”)

- sensor size

1.5 x 8 mm (0.06 x 0.315”)

- cross section

1.77 mm² (0.0027-inch²)

- body

0.01 to 5 BarA
-40°C to +140°C

- sensor

-200°C to +200°C

Accuracy

±10 mbar ≤ 3 BarA @ 130°C; ±15 mbar > 3 BarA @ 130°C; ±15 mbar ≤ 3 BarA @ 23°C; ±30 mbar > 3 BarA @ 23°C; ±0.1°C from -40 to +140°C

Resolution

0.1mbar; 0.01°C

Memory capacity

16,000 data points pressure / 16,000 data points temperature

Memory type

nonvolatile

Sampling rate

1 sec to 24 hours

Transmission

simultaneous: maximum 20 loggers
sequential: unlimited

Internal clock drift

4 sec / 24 h @ 23°C

Mechanical

- material

Peek 1000 - FDA compliant

- enclosure

IP-68

- weight

129 g (4.55 oz)

- diameter

64.50 mm (2.54”)

- height

51.0 mm (2.01”)

Power

2 x Lithium batteries, 0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970,
user replaceable

Battery lifetime

Up to 6 months, based on usage

Antenna

Internal

Wireless range

121 m (400 feet)

Radio frequency

2.4 Ghz ISM band

Access point

USB to computer
Programs and reads unlimited number of loggers

Calibrations

Factory Calibrations:
User Calibrations:

traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
closed loop calibration using XpertLog® software

Standard Calibration Points

hysteresis at 23°C and 130°C:
• 0.05 BarA
• 0.50 BarA
• 1.00 BarA
• 1.50 BarA
• 2.00 BarA
• 2.50 BarA
• 3.00 BarA
• 3.50 BarA
• 4.00 BarA
• 4.50 BarA
• 5.00 BarA

Certifications

EMC:
Health Safety:
RF:

EN 60601-1-2 / 2007
EN 60950-1 / 2006
EN 300440-2 v1.3.1 (2009-03)