Xpertlog® GMPF RF 30H small size real-time temperature data logger designed to fit between 30H vials (-90ºC a +140ºC).

-90ºC to + 140ºC

Applications: Autoclaves, incubators, ovens, on-site steam, stability chambers, warehouses, lyophilisates, freezers, ultra-freezers, refrigerators, cold rooms, cold chain shipments, logistics.

Specs:

· Collects and transmits temperature in real time to XpertLog software.

· ITS-90 calibration coefficients are stored in internal memory.

· Complies with FDA 21 CFR Part 11.

· Network ready: multiple users, centralized, server database, active directory authentication

· Calibration can be done on site.

· Can be activated remotely

Data sheet

Characteristics

GMPF_RF_30H/GMPF_RF_30H_THIN

Sensor type

4 wires / 4 wires thin

- flexible

4 wires: 480 mm (19”) / 4 wires thin: 150 mm (6") or custom length

- sensor size

4 wires: 1.5 x 8 mm (0.06 x 0.315”)
4 wires thin: 1.2 x 2.8 mm (0.05 x 0.11”)

- cross section

4 wires: 1.77 mm² (0.0027-inch²)
4 wires thin: 1.13 mm² (0.00175 inch²)

- body

-90 °C to +140 °C

- sensor

4 wires: -200°C to +200°C
4 wires thin: -90°C to +140°C

Accuracy

4 wires: ±0.1°C from -90 to +140°C
4 wires thin: ±0.1°C from -90 to +140°C

Resolution

0.01°C

Memory capacity

32,000 data points

Memory type

nonvolatile

Sampling rate

1 sec to 24 hours

Transmission

simultaneous: maximum 20 loggers
sequential: unlimited

Internal clock drift

4 sec / 24 h @ 23°C

Mechanical

- material

Peek 1000 - FDA compliant

- enclosure

IP-68

- weight

50 g (1.76 oz)

- diameter

37 mm (1.46”)

- height

62.81 mm (2.47”)

Power

2 x Lithium batteries, 0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970,
user replaceable

Battery lifetime

Up to 6 months, based on usage

Antenna

Internal

Wireless range

121 m (400 feet)

Radio frequency

2.4 Ghz ISM band

Access point

USB to computer
Programs and reads unlimited number of loggers

Calibrations

Factory Calibrations:

User Calibrations:

traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
closed loop calibration using XpertLog® software

Standard Calibration Points

• -80°C
• 0°C
• 140°C

Certifications

EMC:
Health Safety:
RF:

EN 60601-1-2 / 2007
EN 60950-1 / 2006
EN 300440-2 v1.3.1 (2009-03)