-90ºC to + 140ºC
Easy to program and install, with no data loss during the validation process, multiple users can share the same database simultaneously. Factory remote calibration at users' site.
Applications: Autoclaves, incubators, ovens, on-site steam, stability chambers, warehouses, lyophilisates, freezers, ultra-freezers, refrigerators, cold rooms, cold chain shipments, logistics.
Specs:
· Collects and stores temperature information in internal memory
· Retrieve data via docking station using XpertLog software
· ITS-90 calibration coefficients are stored in internal memory
· Complies with FDA 21 CFR Part 11
Network ready: multiple users, centralized, server database, active directory authentication
Data sheet
Characteristics |
GMPL |
|
Sensor
type |
4 wires PT-100 |
|
- rigid |
150 mm (6”) |
|
- customized |
available |
|
- sensor size |
1.5 x 8 mm (0.06 x
0.315”) |
|
- cross section |
1.77 mm² (0.0027-inch²) |
|
- body |
-90 °C to +140 °C |
|
- sensor |
-90 °C to +140 °C |
|
Accuracy |
±0.1°C from -90 to
140°C |
|
Resolution |
0.01°C |
|
Memory
capacity |
32,000 data points |
|
Memory
type |
nonvolatile |
|
Sampling
rate |
1 sec to 24 hours |
|
Internal
clock drift |
4 sec / 24 h @ 23°C |
|
Mechanical |
||
- material |
Peek 1000 - FDA
compliant |
|
- enclosure |
IP-68 |
|
- weight |
50 g (1.76 oz) |
|
- diameter |
40.0 mm (1.57”) |
|
- height |
50.0 mm (1.97”) |
|
Power |
1 x Lithium
batteries, 0.35 g/each Lithium content |
|
Battery
lifetime |
Up to 12 months,
based on usage |
|
Docking
station |
USB to computer |
|
Calibrations |
Factory Calibrations: User Calibrations: |
traceable
NIST/COFRAC - ITS-90 coefficients stored in the internal memory |
Standard
Calibration Points |
• -80°C |
|
Certifications |
EMC: |
EN 60601-1-2 / 2007 |