We import your parts

GMP SW

GMP SW

Standard temperature data logger to capture temperature values within plug washers (-40 ° C to + 140 ° C).

-40ºC to + 140ºC

Easy to program and install, highly reliable (no data loss), high precision (± 0.1 ° C).

Applications: Autoclaves, incubators, ovens, on-site steam, stability chambers, warehouses, lyophilisates, freezers, ultra-freezers, refrigerators, cold rooms, cold chain shipments, logistics.

Specs:

· Collects and stores temperature information in internal memory

· Retrieve data via docking station using XpertLog software

· ITS-90 calibration coefficients are stored in internal memory

· Complies with FDA 21 CFR Part 11

· Network ready: multiple users, centralized, server database, active directory authentication

Data sheet

Characteristics

GMP_SW

Sensor type

4 wires PT-100

- rigid

20 mm (0.8”)

- sensor size

1.5 x 8 mm (0.06 x 0.315”)

- cross section

1.77 mm² (0.0027-inch²)

- body

-40 °C to +140 °C

- sensor

-40 °C to +140 °C

Accuracy

±0.1°C from -40 to 140°C

Resolution

0.01°C

Memory capacity

32,000 data points

Memory type

nonvolatile

Sampling rate

1 sec to 24 hours

Internal clock drift

4 sec / 24 h @ 23°C

Mechanical

- material

Peek 1000 - FDA compliant

- enclosure

IP-68

- weight

91 g (3.21 oz)

- diameter

50.0 mm (1.97”)

- height

40.0 mm (1.57”)

Power

1 x Lithium batteries, 0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970,
user replaceable

Battery lifetime

Up to 12 months, based on usage

Docking station

USB to computer
4, 8, 12 and 20 ports available
Programs and reads unlimited number of loggers

Calibrations

Factory Calibrations:

User Calibrations:

traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
closed loop calibration using XpertLog® software

Standard Calibration Points

• -40°C
• 0°C
• 140°C

Certifications

EMC:
Health Safety:

EN 60601-1-2 / 2007
EN 60950-1 / 2006