We import your parts

GMPF LP

GMPF LP

Low profile standard temperature data logger (35mm high) with long flexible sensor for recording temperature values (-90ºC a +140ºC).

-90ºC to + 140ºC

Easy to program and install, with no data loss during the validation process, multiple users can simultaneously share the same database. Factory remote calibration at users' site.

Applications: Autoclaves, incubators, ovens, on-site steam, stability chambers, warehouses, lyophilisates, freezers, ultra-freezers, refrigerators, cold rooms, cold chain shipments, logistics.

Specs

· Collects and stores temperature information in internal memory

· Retrieve data via docking station using XpertLog software

· ITS-90 calibration coefficients are stored in internal memory

· Complies with FDA 21 CFR Part 11

· Network ready: multiple users, centralized, server database, active directory authentication

· Calibration can be done on site

Data sheet

Characteristics

GMPF-LP

Sensor type

4 wires PT-100

- flexible

480 mm (19”)

- customized

up to 5000 mm (195”)

- sensor size

1.5 x 8 mm (0.06 x 0.315”)

- cross section

1.77 mm² (0.0027-inch²)

- body

-90 °C to +140 °C

- sensor

-200 °C to +200 °C

Accuracy

±0.1°C from -90 to 140°C

Resolution

0.01°C

Memory capacity

32,000 data points

Memory type

nonvolatile

Sampling rate

1 sec to 24 hours

Internal clock drift

4 sec / 24 h @ 23°C

Mechanical

- material

Peek 1000 - FDA compliant

- enclosure

IP-68

- weight

44 g (1.55 oz)

- diameter

40.0 mm (1.57”)

- height

35.0 mm (1.38”)

Power

1 x Lithium batteries, 0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970,
user replaceable

Battery lifetime

Up to 12 months, based on usage

Docking station

USB to computer
4, 8, 12 and 20 ports available
Programs and reads unlimited number of loggers

Calibrations

Factory Calibrations:

User Calibrations:

traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
closed loop calibration using XpertLog® software

Standard Calibration Points

• -80°C
• 0°C
• 140°C

Certifications

EMC:
Health Safety:

EN 60601-1-2 / 2007