We import your parts

XPERTVAL

XPERTVAL

60 channel (3x20) thermocouple based system for thermal mapping and validation (up to + 1200 ° C).

up to + 1200 ° C

Specs:

· System based on conventional Tc for validation and thermal mapping

· 60 (up to 120) channels simultaneously

· Rugged data logger used in harsh industrial environments, real-time data display

· High precision Tc after precalibration (± 0.2 ° C)

· Complies with FDA 21 CFR Part 11

· Network ready: multiple users, centralized, server database, active directory authentication

Data sheet

Characteristics

XpertVal

Sensors

Thermocouples

10 m (33 feet)

custom length

Sensor type

All type of thermocouples, most used: T, J, K

- teflon:

-200°C to +200°C

- kapton:

-200°C to +400°C

- other:

up to +1200°C

Accuracy

±0.2°C after pre-calibration

Memory capacity

Internal memory: 20 Mb
Backup memory: USB storage device compatible

Resolution

0.01°C

TC Tip

Teflon: Teflon tip to prevent moisture on the wire
Kapton: no tip

Sampling rate

120 channels - 5 sec
60 channels - 5 sec
20 channels - 2 sec
12 channels - 1 sec

Software

21 CFR Part 11 compliant

Module

Communication USB XpertVal to PC
LAN
2.4 GHz, 802.11b/g/n
Close loop calibration Required when using thermocouples for thermal validation
System uses temperature block and reference probe, software-controlled reports generated automatically by the software

Reference probes

RTD, Fluke 1502, Lives XpertRTD

Mechanical

Dry blocks

Lives: PTC 125: -90°C to +125°C / RTC 157: -45°C to +155°C / RTC 159: -100°C to +155°C
RTC 700: +23°C to +700°C
Fluke 9190: -90°C to +140°C; Fluke 9170: -45°C to +140°C; Fluke 9172: +35°C to +425°C

Factory:

NIST / COFRAC

Users:

close loop calibration using XpertLog® software